Catalog Number 381312 |
Device Problem
Detachment of Device or Device Component (2907)
|
Patient Problem
Foreign Body In Patient (2687)
|
Event Date 02/21/2024 |
Event Type
Injury
|
Manufacturer Narrative
|
H.3.If a device evaluation and/or device history review is completed, a supplemental report will be filed.
|
|
Event Description
|
It was reported that bd insyte-w winged yel 24ga x 0.75in catheter broke the following information was provided by the initial reporter: the patient was 1 year old and 6 months old and placed an insyte indwelling needle on sunday, and the normal infusion was made on february 20, and the abnormality was found on the morning of the 21st, only a small half was cut outside, and the superficial subcutaneous vein was found to have 15mm imaging around the eye of the needle on the back of the left hand by imaging examination.At present, the child was arrested by surgery, but the child was not caught by surgery because he was too young.The hospital asked to find a hospital that could perform surgery to arrest successful cases to assist in the operation.
|
|
Event Description
|
The patient was 1 year old and 6 months old and placed an insyte indwelling needle on sunday, and the normal infusion was made on february 2 0, and the abnormality was found on the morning of the 21st, only a small half was cut outside, and the superficial subcutaneous vein was found to have 15mm imaging around the eye of the needle on the back of the left hand by imaging examination.At present, the child was arrested by surgery, but the child was not caught by surgery because he was too young.The hospital asked to find a hospital that could perform surgery to arrest successful cases to assist in the operation.
|
|
Manufacturer Narrative
|
Based on device history record review, no abnormality was observed during the production run.The catheter separation test results for outgoing inspection were within specifications.From the returned photos, catheter tubing broke off at mid of catheter tubing.The assembly process was reviewed, and the only station that is in contact with the catheter tubing is the rubber pads at the flaring station.Given the elastic nature of the rubber pads and there are no sharp or hard edges on the pads, this station is unlikely to cause damage on the catheter tubing to result in reported defect.As current control for catheter damage, there is an outgoing functional test in place to check for catheter to adapter pull force.There is also an outgoing visual inspection and in-process visual inspection in place to check for catheter damage.As the actual sample not returned, actual root cause could not be established.
|
|
Search Alerts/Recalls
|
|