WILSON-COOK MEDICAL INC ACROBAT® 2 CALIBRATED TIP WIRE GUIDE; OCY ENDOSCOPIC GUIDEWIRE, GASTROENTEROLOGY-UROLOGY
|
Back to Search Results |
|
Catalog Number AWG2-25-450 |
Device Problem
Peeled/Delaminated (1454)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/16/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
Investigation evaluation: a product evaluation was performed only by the picture provided with this report because the product said to be involved was not provided to cook for evaluation.A photo of the lot number was not provided.Our evaluation of the photo provided confirmed the report.The coating of the wire guide has split exposing the core wire.The two (2) band marker, which is located approximately 10 cm from the distal tip, is visible near the damage.Coating damage is noted on both sides of the two (2) band marker, though one side is a small area and the other is a long split segment.However, due to the condition of the device we are unable to determine if the majority of damage is proximal or distal to the two (2) band marker.A product discrepancy or anomaly that could have contributed to this reported occurrence was not observed.A review of the device history record could not be conducted because the lot number was not provided.Investigation conclusion: our evaluation of the image provided confirmed the report of coating damage.The user states that pressure was being applied to the wire guide by the endoscope elevator, this is a likely cause of the coating damage observed.If additional pressure is applied to the wire guide and/or accessory device(s) while moving the wire guide inside the accessory device(s), this could contribute to wire guide coating damage.Prior to distribution, all acrobat® 2 calibrated tip wire guides are subjected to a visual inspection and functional testing to ensure device integrity.Corrective action: a review of the complaint history was conducted.The likelihood of occurrence is considered rare.Corrective action is not warranted at this time based on the quality engineering risk assessment.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.
|
|
Event Description
|
During an endoscopic retrograde cholangiopancreatography (ercp) procedure, the physician used a cook acrobat® 2 calibrated tip wire guide.It was reported that the physician performed a sphincterotomy, and about 3-4 balloon sweeps, he noticed the wire looking like it had ¿split¿.Procedure went ok, and no adverse events took place.Physician did not complain about the wire, and believed that it may have been because of the elevator in the scope being pushed up on the wire as it was pulled out.An image of the device was provided to show the coating damage.A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
|
|
Search Alerts/Recalls
|
|
|