• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) ASSURITY RF DR; Pulse generator, permanent, implantable

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) ASSURITY RF DR; Pulse generator, permanent, implantable Back to Search Results
Model Number PM2240
Device Problem Defective Device (2588)
Patient Problem Insufficient Information (4580)
Event Date 03/13/2024
Event Type  malfunction  
Manufacturer Narrative
No complaint was received with the return of the device.Failure event observed during analysis.The device was received with normal telemetry and output.Visual inspection of the header attachment area detected a bonding anomaly.A device history record (dhr) review was performed and all required manufacturing processes and inspections steps were confirmed to be completed per the requirements.The device met specifications prior to leaving abbott manufacturing facilities.The device was cut open to enable further testing and longevity assessment found the device was operating at the elective replacement indicator (eri) voltage consistent with normal usage.Hybrid circuitry was tested, and the results indicated normal current drain.A manufacturing anomaly may have occurred, which resulted in the header bonding anomaly.The device is included in the assurity and endurity pacemakers header anomaly advisory issued by abbott on 15 march 2021.
 
Event Description
This report is to advise of an event observed during analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ASSURITY RF DR
Type of Device
Pulse generator, permanent, implantable
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key18909705
MDR Text Key337739467
Report Number2017865-2024-35049
Device Sequence Number1
Product Code NVZ
UDI-Device Identifier05414734507073
UDI-Public05414734507073
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P880086
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other
Reporter Occupation Non-Healthcare Professional
Remedial Action Recall
Type of Report Initial
Report Date 03/14/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date07/31/2017
Device Model NumberPM2240
Device Lot NumberA000016419
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/05/2024
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 03/13/2024
Initial Date FDA Received03/14/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/12/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberZ-1530-2021
Patient Sequence Number1
-
-