• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHYSIO-CONTROL, INC. - 3015876 LIFEPAK(R) 1000 DEFIBRILLATOR; AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHYSIO-CONTROL, INC. - 3015876 LIFEPAK(R) 1000 DEFIBRILLATOR; AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE) Back to Search Results
Model Number 1000
Device Problem Failure to Sense (1559)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/19/2024
Event Type  malfunction  
Event Description
The customer contacted stryker to report that their device was unable to detect defibrillation electrodes.In this state, the device may not be able to provide defibrillation therapy, if it were needed.This issue is patient related; however there was no adverse patient outcome reported.
 
Manufacturer Narrative
Stryker contacted the customer to request additional information on the patient.No response has been received from the customer.Patient fields in which information is not provided were intentionally left blank. a stryker service representative performed an initial evaluation of the customer¿s device and was not able to verify the reported issue.Proper device operation was observed through functional and performance testing.The device was subsequently returned to the customer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEPAK(R) 1000 DEFIBRILLATOR
Type of Device
AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)
Manufacturer (Section D)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer (Section G)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer Contact
brian blakeslee
11811 willows road ne
redmond, WA 98052
4258674000
MDR Report Key18910247
MDR Text Key337739249
Report Number0003015876-2024-00546
Device Sequence Number1
Product Code MKJ
UDI-Device Identifier00883873815458
UDI-Public00883873815458
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K122600
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 03/15/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number1000
Device Catalogue Number99425-000023
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/19/2024
Initial Date Manufacturer Received 02/19/2024
Initial Date FDA Received03/15/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/16/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-