Catalog Number 451570000
Device Problem Entrapment of Device (1212)
Patient Problem Foreign Body In Patient (2687)
Event Date 03/07/2024
Event Type  Injury  
Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event Description
It was reported that the saw blade detached from saw hand piece while cutting femur, saw blade stayed embedded in the femur.Black liquid was noted around where the saw blade attaches to the saw hand piece.Was surgery delayed due to the reported event? yes.If yes, number of minutes: 5.
 
Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event Description
Additional information was received and stated that there were no reports of any patient consequences.
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: according to the information received, ¿saw blade detached from saw hand piece while cutting femur, saw blade stayed embedded in the femur.Black liquid was noted around where the saw blade attaches to the saw hand piece.¿ the product was not returned to depuy synthes, however photos were provided for review.Review of the photographic evidence reveled a velys saw blade unassembled from the velys saw handpiece; however, no evidence of the device being embedded to patient was provided.In addition, a black substance was observed on the proximal section of the blade.The origin of the substance is unknown, but it is suspected to be related to the handpiece.It is worth mentioning, that photos of the product code and serial number of the handpiece were provided.The overall complaint was not confirmed as the photographs provided contained insufficient evidence of the reported event.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VELYS SAW BLADE
Type of Device
INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
Manufacturer (Section G)
DEPUY SYNTHES PRODUCTS LLC
4500 riverside drive
palm beach gardens FL
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581-0988
3035526892
MDR Report Key18912725
MDR Text Key337763177
Report Number1818910-2024-05826
Device Sequence Number1
Product Code OLO
UDI-Device Identifier10603295519140
UDI-Public10603295519140
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
K202769
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date 03/15/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue Number451570000
Device Lot NumberVE200024
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/06/2024
Initial Date FDA Received03/15/2024
Supplement Dates Manufacturer Received05/13/2024
09/29/2015
Supplement Dates FDA Received05/17/2024
05/30/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/15/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-