ZIMMER BIOMET, INC. TPRLC 133 MP TYPE1 PPS HO 10.0; PROSTHETIC, HIP Back to Search Results
Model Number N/A
Device Problem Manufacturing, Packaging or Shipping Problem (2975)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/27/2024
Event Type  malfunction  
Manufacturer Narrative
(b)(4) visual evaluation of the returned product identified damage to the sterile packaging (blister).Sterility has been compromised.Review of the device history records identified no deviations or anomalies during manufacturing related to the reported event.Medical records were not provided.The condition of the device when it left zimmer biomet is considered conforming to specification.The root cause of the reported event can be attributed to transit damage and a packaging design issue.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
It was reported that the inner packaging was damaged.There was no discernible damage to exterior packaging.There was no patient impact or delay reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TPRLC 133 MP TYPE1 PPS HO 10.0
Type of Device
PROSTHETIC, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key18913065
MDR Text Key337768595
Report Number0001825034-2024-00695
Device Sequence Number1
Product Code KWA
UDI-Device Identifier00880304513433
UDI-Public(01)00880304513433(17)330317(10)7459881
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K110400
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 03/12/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number51-107100
Device Lot Number7459881
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/04/2024
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 02/27/2024
Initial Date FDA Received03/15/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/17/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Removal/Correction NumberN/A
Patient Sequence Number1
Patient SexMale
-
-