• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) TENDRIL STS; NO MATCH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) TENDRIL STS; NO MATCH Back to Search Results
Model Number 2088TC/46
Device Problems Capturing Problem (2891); Device Sensing Problem (2917); Device Dislodged or Dislocated (2923)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  Injury  
Event Description
It was reported that a patient presented to clinic for follow up.Device interrogation revealed high capture threshold and low p-waves on the atrial lead.Chest x-ray was performed and revealed atrial lead dislodgement.The physician elected to explant and replace the atrial lead.The patient condition was stable.
 
Manufacturer Narrative
The reported events were lead dislodgement, unacceptable threshold, and p-wave amp variation.Visual examination of the lead found the helix was extended and clogged with blood / tissue.After cleaning and applying torque directly to the connector pin, the helix could be retracted and extended.The full helix extension length was measured within specification.The reported events unacceptable threshold and p-wave amp variation were not confirmed.Electrical testing did not find any indication of conductor fractures or internal shorts.Visual and x-ray inspections of the lead did not find any anomalies except for procedural damage.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TENDRIL STS
Type of Device
NO MATCH
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key18913638
MDR Text Key337776647
Report Number2017865-2024-35094
Device Sequence Number1
Product Code NVN
UDI-Device Identifier05414734502870
UDI-Public05414734502870
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
P960013
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/28/2023
Device Model Number2088TC/46
Device Lot NumberP000101411
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 02/26/2024
Initial Date FDA Received03/15/2024
Supplement Dates Manufacturer Received04/11/2024
Supplement Dates FDA Received04/26/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/25/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
ASSURITY PM; TENDRIL RV LEAD
Patient Outcome(s) Required Intervention;
Patient Age62 YR
Patient SexMale
-
-