The reported events were lead dislodgement, unacceptable threshold, and p-wave amp variation.Visual examination of the lead found the helix was extended and clogged with blood / tissue.After cleaning and applying torque directly to the connector pin, the helix could be retracted and extended.The full helix extension length was measured within specification.The reported events unacceptable threshold and p-wave amp variation were not confirmed.Electrical testing did not find any indication of conductor fractures or internal shorts.Visual and x-ray inspections of the lead did not find any anomalies except for procedural damage.
|