• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 SURESTEP¿ FOLEY TRAY SYSTEM BARD® LUBRI-SIL® FOLEY CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 SURESTEP¿ FOLEY TRAY SYSTEM BARD® LUBRI-SIL® FOLEY CATHETER Back to Search Results
Catalog Number A947316
Device Problem Gel Leak (1267)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/13/2024
Event Type  malfunction  
Event Description
It was reported that lubricating jelly syringe of foley tray was present but empty.Lube provided from another source outside of product.Product used and then disposed of in the proper fashion.Unable to provide defective product.
 
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Manufacturer Narrative
The reported event was inconclusive because no sample was returned.A potential root cause for this failure could be due to "improper grommet fit within barrel resulting in blow by/spillage of lube".A review of the device history record did not show any problems or conditions that would have contributed to the reported issue.A labeling review is not required because labeling could not have prevented the reported issue.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text: the device was not returned.
 
Event Description
It was reported that lubricating jelly syringe of foley tray was present but empty.Lube provided from another source outside of product.Product used and then disposed of in the proper fashion.Unable to provide defective product.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SURESTEP¿ FOLEY TRAY SYSTEM BARD® LUBRI-SIL® FOLEY CATHETER
Type of Device
FOLEY TRAY
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
xeeroy rada
8195 industrial blvd
covington 30014
7707846100
MDR Report Key18913977
MDR Text Key337781326
Report Number1018233-2024-01363
Device Sequence Number1
Product Code EZL
UDI-Device Identifier00801741074141
UDI-Public(01)00801741074141
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K040504
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/05/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberA947316
Device Lot NumberNGHU2141
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/26/2024
Initial Date FDA Received03/15/2024
Supplement Dates Manufacturer Received04/05/2024
Supplement Dates FDA Received04/10/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/18/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-