ABBOTT IRELAND DIAGNOSTICS DIVISION ARCHITECT FREE T4 REAGENT KIT; RADIOIMMUNOASSAY, FREE THYROXINE Back to Search Results
Catalog Number 07K65-77
Device Problem Low Test Results (2458)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/06/2024
Event Type  malfunction  
Manufacturer Narrative
An evaluation is in process.A follow-up report will be submitted when the evaluation is complete.All available patient information has been included.No additional patient details are available.
 
Event Description
The customer observed falsely depressed architect free t4 results for three patients that were not consistent with patient history.The following data was provided (unit: pmol/l): patient 1, 87-year-old male, diagnosed with type 2 diabetes with multiple complications; initial result 7.45, repeated 9.32.Patient 2, 63-year-old woman diagnosed with type 2 diabetes; initial result 7.54, repeated 9.56.Patient 3, 24-year-old male was diagnosed with hypokalemia; initial result 8.79, repeated 11.13.(reference range 9.01-19.05).No impact to patient management was reported.
 
Manufacturer Narrative
Correction found in section d4 lot number.A review of customer data aligned with customer¿s issue and no additional issues were identified.The ticket search by lot indicates that the reagent lot shows normal complaint activity.Device history review did not identify any issues associated with the customer¿s observation.A review of ticket trending did not identify any related trend regarding commonalities for complaint lot number and issue.Retained analysis of architect free t4 reagent lot 50481ud00.Testing met acceptance criteria and the product is performing as expected.Labeling review concludes that the issue is adequately addressed.Based on the investigation, no systemic issue or deficiency of the architect free t4 assay was identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARCHITECT FREE T4 REAGENT KIT
Type of Device
RADIOIMMUNOASSAY, FREE THYROXINE
Manufacturer (Section D)
ABBOTT IRELAND DIAGNOSTICS DIVISION
lisnamuck
co. longford
longford N39E9 32
EI  N39E932
Manufacturer (Section G)
ABBOTT IRELAND DIAGNOSTICS DIVISION
lisnamuck
co. longford
longford N39E9 32
EI   N39E932
Manufacturer Contact
nicole jenne
max-planck-ring 2
post market surveillance
wiesbaden 65205
GM   65205
6122582960
MDR Report Key18915329
MDR Text Key337798841
Report Number3005094123-2024-00125
Device Sequence Number1
Product Code CEC
UDI-Device Identifier00380740173678
UDI-Public(01)00380740173678(17)240306(10)50481UD02
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K173122
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 05/13/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/06/2024
Device Catalogue Number07K65-77
Device Lot Number50481UD02
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/06/2024
Initial Date FDA Received03/15/2024
Supplement Dates Manufacturer Received04/27/2024
Supplement Dates FDA Received05/13/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/26/2023
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
ARC I2000SR INST, 03M74-02, (B)(6) ; ARC I2000SR INST, 03M74-02, (B)(6)
-
-