It was reported that the patient with this pacemaker was evaluated in clinic for a routine follow up with the physician.At the moment of the evaluation, the patient stated feeling dizzy.An electrocardiogram was performed, and it was noted a heart rate of 29 beats per minute (bpm) and no signs of pacing.The device was not able to be interrogated.A magnet was applied, and no response was observed.A device changeout procedure was performed due to premature battery depletion and it was explanted and replaced.It was noted the battery of this device depleted at less than 4 years after it was implanted.No additional adverse patient effects were reported.It is expected to receive this device for analysis.
|