RESPIRONICS, INC. RESPIRONICS; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
|
Back to Search Results |
|
Model Number V60 V60PLUS VENTILATOR |
Device Problem
Failure to Sense (1559)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/13/2024 |
Event Type
malfunction
|
Event Description
|
Philips received a complaint by the customer on the v60, indicating that a 1202 "proximal pressure line disconnect" alarm sounded.There was no patient involvement at the time the issue was discovered the issues were found during periodical device checking.There was no reported delay in therapy.The customer called technical support to report that a 1202 "proximal pressure line disconnect" alarm sounded.The manufacturer's product support engineer (pse) evaluated the device, and although the pse confirmed the 1202 error code in the event log, the pse could not actually duplicate the reported issue as no malfunction was found.The device passed the required performance verification tests per philips standard and was returned to service.The investigation concludes that no further action is required at this time.If additional information is received, the complaint file will be reopened.
|
|
Manufacturer Narrative
|
E1: (b)(6).
|
|