• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS, INC. RESPIRONICS; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS, INC. RESPIRONICS; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE Back to Search Results
Model Number V60 V60PLUS VENTILATOR
Device Problem Failure to Sense (1559)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/13/2024
Event Type  malfunction  
Event Description
Philips received a complaint by the customer on the v60, indicating that a 1202 "proximal pressure line disconnect" alarm sounded.There was no patient involvement at the time the issue was discovered the issues were found during periodical device checking.There was no reported delay in therapy.The customer called technical support to report that a 1202 "proximal pressure line disconnect" alarm sounded.The manufacturer's product support engineer (pse) evaluated the device, and although the pse confirmed the 1202 error code in the event log, the pse could not actually duplicate the reported issue as no malfunction was found.The device passed the required performance verification tests per philips standard and was returned to service.The investigation concludes that no further action is required at this time.If additional information is received, the complaint file will be reopened.
 
Manufacturer Narrative
E1: (b)(6).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESPIRONICS
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
kimberly shelly
1001 murry ridge lane
murrysville, PA 15668
7247330200
MDR Report Key18920254
MDR Text Key337848706
Report Number2518422-2024-14063
Device Sequence Number1
Product Code MNT
UDI-Device Identifier00884838033832
UDI-Public(01)00884838033832
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K102985
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Other
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberV60 V60PLUS VENTILATOR
Device Catalogue Number1076709
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 03/13/2024
Initial Date FDA Received03/18/2024
Date Device Manufactured08/13/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? Yes
Type of Device Usage Reuse
Patient Sequence Number1
-
-