• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL NAVITOR TRANSCATHETER AORTIC VALVE; AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIVERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL NAVITOR TRANSCATHETER AORTIC VALVE; AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIVERED Back to Search Results
Catalog Number NVTR-29
Device Problem Material Deformation (2976)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/22/2024
Event Type  malfunction  
Manufacturer Narrative
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Event Description
It was reported that on (b)(6) 2024, a 29 navitor valve was chosen for implant, using large flexnav delivery system.During implant, the stent frame was observed to be bent in fluoroscopy.Nothing unusual with the device during preparation.It was decided to remove the device.A replacement 29mm navitor valve was successfully implanted, using a replacement large flexnav delivery system.The patient remained hemodynamically stable throughout the procedure.There was no clinically significant delay in procedure and no adverse patient effects.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NAVITOR TRANSCATHETER AORTIC VALVE
Type of Device
AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIVERED
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ST. JUDE MEDICAL #3014918977
edificio #44 calle 0, ave. 2
el coyol alajuela 1897- 4050
CS   1897-4050
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key18922419
MDR Text Key337882785
Report Number2135147-2024-01187
Device Sequence Number1
Product Code NPT
UDI-Device Identifier05415067031600
UDI-Public05415067031600
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
P190023
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 03/18/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberNVTR-29
Device Lot Number9133938
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/22/2024
Initial Date FDA Received03/18/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/20/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment