Model Number LF-DP
Device Problem Peeled/Delaminated (1454)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
It was observed during device evaluation that the coating on the connecting tube of the intubation fiberscope was peeling.There was no patient involvement.
 
Manufacturer Narrative
The device was returned, and the evaluation found no additional reportable malfunctions aside from the allegation reported in b5.Should additional relevant information become available, a supplemental report will be submitted.
 
Manufacturer Narrative
This supplemental report is being submitted to provide the results of the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 19 years since the subject device was manufactured.Based on the results of the investigation, the coating at the insertion tube was peeled off, likely due to the stress of repeated use, external factors, or handling of the device.However, a definitive root cause could not be determined.The instructions for use states the inspection method associated with the event in "chapter 3 preparation and inspection 3.2 inspection of the endoscope.¿ olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRACHEAL INTUBATION FIBERSCOPE
Type of Device
TRACHEAL INTUBATION FIBERSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18923865
MDR Text Key337903124
Report Number9610595-2024-05716
Device Sequence Number1
Product Code EOQ
UDI-Device Identifier04953170340192
UDI-Public04953170340192
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K981543
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 04/23/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberLF-DP
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/03/2024
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 03/05/2024
Initial Date FDA Received03/18/2024
Supplement Dates Manufacturer Received04/17/2024
Supplement Dates FDA Received04/23/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/12/2004
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-