Model Number L311 |
Device Problems
Premature Discharge of Battery (1057); Battery Problem (2885)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/06/2024 |
Event Type
Injury
|
Event Description
|
It was reported that this device was suspected of exhibiting premature battery depletion behavior.It was noted that the patient has atrial flutter.The device was programmed to a dual chamber (dddr) mode to pace and sense both the atrium and the ventricle.The patient was almost 100% right ventricular paced.The rv auto output was set to auto capture @ 5.0v @ 0.4ms and the battery had 5 months remaining.The pacemaker was adjusted to a single chamber mode (vvir) with a fixed rv output, only increasing the battery longevity to 7 months.Subsequently, this device was explanted and replaced.No additional adverse patient effects were reported.This device is expected for return.
|
|
Manufacturer Narrative
|
As no further information concerning this report is expected, our investigation is complete.This investigation will be updated should further information be provided.
|
|
Event Description
|
It was reported that this device was suspected of exhibiting premature battery depletion behavior.It was noted that the patient has atrial flutter.The device was programmed to a dual chamber (dddr) mode to pace and sense both the atrium and the ventricle.The patient was almost 100% right ventricular paced.The rv auto output was set to auto capture @ 5.0v @ 0.4ms and the battery had 5 months remaining.The pacemaker was adjusted to a single chamber mode (vvir) with a fixed rv output, only increasing the battery longevity to 7 months.Subsequently, this device was explanted and replaced.No additional adverse patient effects were reported.This device is expected for return.This device was not returned for analysis.Despite several requests to obtain product return, no further information was provided.
|
|
Search Alerts/Recalls
|