AIZU OLYMPUS CO., LTD. GASTROINTESTINAL VIDEOSCOPE Back to Search Results
Model Number GIF-EZ1500
Device Problem Device Reprocessing Problem (1091)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
It was observed that during the device evaluation, the gastrointestinal videoscope nozzle was clogged with surgical glue.There were no reports of patient involvement.
 
Manufacturer Narrative
The device was returned, and the evaluation found that due to clogging of nozzle, water supply volume did not meet the standard value.Should additional relevant information become available, a supplemental report will be submitted.
 
Manufacturer Narrative
This supplemental report is being submitted to provide the results of the legal manufacturer's final investigation.Corrected fields: d9 additional information added to field: h6 please note h3 was inadvertently selected and cannot be removed, disregard this selection as it was already selected on the initial.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the foreign material clogging the nozzle was believed to be surgical glue.It is likely the foreign material remained in the device because reprocessing could not be conducted properly due to a leak from the biopsy channel.However, a definitive root cause could not be determined.The suggested event may be detected and prevented by handling device in accordance with the following ifu.Ifu: gif-ez1500 operation manual chapter 3 preparation and inspection states how to detect the event.Ifu: gif-ez1500 reprocessing manual chapter 5 reprocessing the endoscope states how to prevent the event.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GASTROINTESTINAL VIDEOSCOPE
Type of Device
GASTROINTESTINAL VIDEOSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18924682
MDR Text Key337914127
Report Number9610595-2024-05728
Device Sequence Number1
Product Code FDS
Combination Product (y/n)N
Reporter Country CodeAU
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 04/15/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberGIF-EZ1500
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/07/2024
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 03/12/2024
Initial Date FDA Received03/18/2024
Supplement Dates Manufacturer Received04/02/2024
Supplement Dates FDA Received04/15/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/02/2023
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-