• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS COAGUCHEK XS PT TEST PST; PROTHROMBIN TIME TEST STRIPS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS COAGUCHEK XS PT TEST PST; PROTHROMBIN TIME TEST STRIPS Back to Search Results
Catalog Number 07762798003
Device Problems Low Test Results (2458); Non Reproducible Results (4029)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/26/2024
Event Type  malfunction  
Event Description
There was an allegation of questionable results from the coaguchek inrange meter.The result from the meter was 2.3 inr.One hour later, the clinic's result using a coaguchek meter was 3.5 inr.The therapeutic range was 3.0-3.5 inr and the patient usually tests every second day.
 
Manufacturer Narrative
The coaguchek inrange serial number was (b)(6).The meter and strips have been requested for return.The product has not been received at this time.If the product is returned in the future, a follow-up report will be submitted.On a regular basis, coaguchek strips of lots currently valid in the market are tested as part of routine retention testing and results have passed the internal inspection.E3 - occupation was patient/consumer.
 
Manufacturer Narrative
The returned meter and strips were provided for investigation where they were tested using a high-level control sample.Testing results (qc range = 2.4 ¿ 3.6 inr): qc 1: 2.9 inr.Qc 2: 2.9 inr.Qc 3: 2.9 inr.The obtained qc values were in the allowed range of the used combination strip lot - qc lot.All measurements were without error messages.Medwatch fields d4 and d9 were updated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COAGUCHEK XS PT TEST PST
Type of Device
PROTHROMBIN TIME TEST STRIPS
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhofer strasse 116
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
amy nelson
9115 hague road
indianapolis, IN 46250-0457
MDR Report Key18924712
MDR Text Key337914618
Report Number1823260-2024-00793
Device Sequence Number1
Product Code GJS
Combination Product (y/n)Y
Reporter Country CodeGM
PMA/PMN Number
K170960
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Consumer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 03/28/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number07762798003
Device Lot Number73986119
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 02/27/2024
Initial Date FDA Received03/18/2024
Supplement Dates Manufacturer Received03/26/2024
Supplement Dates FDA Received03/28/2024
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
MARCUMAR
Patient SexFemale
-
-