Model Number V60
Device Problem Failure to Sense (1559)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/01/2024
Event Type  malfunction  
Event Description
Philips received a complaint on the v60 ventilator indicating that there was a proximal pressure sensor auto zero failed alarm.The device was reported to be outside of use at the time of the reported problem.There was no patient or user harm reported.The remote service engineer (rse) informed the customer to perform the following troubleshooting steps in this order: verify the pin alignment between the solenoids and data acquisition (da) printed circuit board assembly (pcba), replace the 3rd and 4th proximal autozero solenoids, and then replace the da pcba.
 
Manufacturer Narrative
In a good faith effort (gfe) response received from the customer, it was stated that the customer did not order any replacement part following the remote service because it was found that the solenoid pins were not lined up.Following the reseating of the data acquisition board, the device was put back into use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESPIRONICS
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
kimberly shelly
1001 murry ridge lane
murrysville, PA 15668
7247330200
MDR Report Key18925010
MDR Text Key337918637
Report Number2518422-2024-14212
Device Sequence Number1
Product Code MNT
UDI-Device Identifier00884838020054
UDI-Public00884838020054
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102985
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberV60
Device Catalogue Number1053617
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/01/2024
Initial Date FDA Received03/18/2024
Supplement Dates Manufacturer Received03/25/2024
Supplement Dates FDA Received03/28/2024
Date Device Manufactured11/30/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-