MAUDE Adverse Event Report: BECTON DICKINSON CARIBE LTD. BD BACTEC¿ PLUS AEROBIC/F CULTURE VIALS (PLASTIC); SYSTEM, BLOOD CULTURING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON CARIBE LTD. BD BACTEC¿ PLUS AEROBIC/F CULTURE VIALS (PLASTIC); SYSTEM, BLOOD CULTURING Back to Search Results
Catalog Number 442023
Device Problem False Positive Result (1227)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/11/2024
Event Type  malfunction  
Event Description
Report 7 of 8.It was reported that while using bd bactec¿ plus aerobic/f culture vials (plastic) that there was a false positive for candia tropical on panel bcid2.No patient impact reported.
 
Manufacturer Narrative
Pma / 510(k)# k113558, k222591.Investigation summary: catalog: 442023.Batch no.: 3110054.Customer reported a molecular false positive result.Neither photos nor returned good samples were received.Bd was unable to reproduce the customer¿s experience with the bactec product based on our internal procedures and the intended use of the product.Retention samples were visually inspected, tested for viable contamination by sub-culturing on tsa, chocolate, sabouraud and schaedler agars plates, voltage output and gram stain.All results were satisfactory.Batch history record review did not identify any evidence for which the customer submitted the complaint.A complaint history review was conducted, and batch has been previously investigated for the reported defect.No trend identified for batch.Complaint is unconfirmed based on retention samples and batch history record review results.
 
Event Description
Report 7 of 8.It was reported that while using bd bactec¿ plus aerobic/f culture vials (plastic) that there was a false positive for candia tropical on panel bcid2.No patient impact reported.
 
Manufacturer Narrative
H.6 investigation summary.Catalog: 442023.Batch no.: 3110054.Customer reported a molecular false positive result.Neither photos nor returned good samples were received.Bd was unable to reproduce the customer¿s experience with the bactec product based on our internal procedures and the intended use of the product.Retention samples were visually inspected, tested for viable contamination by sub-culturing on tsa, chocolate, sabouraud and schaedler agars plates, voltage output and gram stain.All results were satisfactory.Batch history record review did not identify any evidence for which the customer submitted the complaint.A complaint history review was conducted, and batch has been previously investigated for the reported defect.No trend identified for batch.Complaint is unconfirmed based on retention samples and batch history record review results.No corrective actions were required.A cross functional team continually monitors all product complaints for trends and determines if any additional actions are necessary beyond the current investigational process.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD BACTEC¿ PLUS AEROBIC/F CULTURE VIALS (PLASTIC)
Type of Device
SYSTEM, BLOOD CULTURING
Manufacturer (Section D)
BECTON DICKINSON CARIBE LTD.
vicks drive
lot no. 6
cayey PR
Manufacturer (Section G)
BECTON DICKINSON CARIBE LTD.
vicks drive
lot no. 6
cayey PR
Manufacturer Contact
jo doyka
7 loveton circle
sparks, MD 21152
4103164000
MDR Report Key18926547
MDR Text Key337929306
Report Number2647876-2024-00052
Device Sequence Number1
Product Code MDB
UDI-Device Identifier00382904420239
UDI-Public(01)00382904420239
Combination Product (y/n)N
Reporter Country CodeCI
PMA/PMN Number
SEE H.10
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 03/27/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date01/22/2024
Device Catalogue Number442023
Device Lot Number3110054
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/01/2024
Initial Date FDA Received03/18/2024
Supplement Dates Manufacturer Received03/13/2024
Supplement Dates FDA Received04/09/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/20/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-