The reported field event of inappropriate therapy, sensing issue and premature battery depletion was not confirmed.As received, the device was at end of service (eos) upon receipt.Review of the device data showed that the device delivered many consecutive hv therapies in a short period of time.The hv therapies were delivered appropriately, and the device acted normally given its programmed settings.A longevity calculation was performed and found the battery depletion was normal based on the device usage.Telemetry, impedance, sensing, pacing, capacitor maintenance, and high voltage (hv) output functions of the device were tested and found to be normal.No anomalies were found.
|