Intuitive surgical, inc.(isi) has not received the instrument for evaluation.Therefore, the root cause of the customer reported failure mode has not been determined.A follow-up mdr will be submitted if the product is returned and evaluated and/ or if additional information is received.
|
Intuitive surgical, inc.(isi) received a da vinci product to perform failure analysis evaluation.The fenestrated bipolar forceps instrument was analyzed and failure analysis investigations confirmed and replicated the customer-reported complaint.The instrument was found to have a broken main tube at the distal tip.No material was found missing.Components adjacent to this broken main tube did not show damage.An additional observation that was not reported by the site was that the proximal clevis of the instrument was dislodged.The dislodged proximal clevis was attached to the main tube.Additionally, isi followed up with the initial reporter and obtained the following additional information: the instrument was inspected prior to use, as usual, after unwrapping from the sterile pack, and there was no damages or anything out of the ordinary noted.The reason the instrument and cannula were removed together was that the wrist could not be straightened due to physical damage (e.G., broken main tube).The doctor did not remember if the port incision was increased to remove the instrument and cannula together.It was confirmed that the incident involved a fragment falling inside the patient's anatomy, which was retrieved during the same procedure.The fragment was retrieved successfully, and it was not necessary to retrieve it during another surgical procedure.There was no collision with any other instrument or tool during the procedure.
|