• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: FARAPULSE, INC. FARAWAVE PULSED FIELD ABLATION CATHETER, 35MM; CARDIAC IRREVERSIBLE ELEC TROPORATION SYSTEM CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

FARAPULSE, INC. FARAWAVE PULSED FIELD ABLATION CATHETER, 35MM; CARDIAC IRREVERSIBLE ELEC TROPORATION SYSTEM CATHETER Back to Search Results
Lot Number 0008020937
Device Problem Improper Flow or Infusion (2954)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/29/2024
Event Type  malfunction  
Manufacturer Narrative
The device has not been received for analysis.Upon receipt and completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.
 
Event Description
It was reported that during an irreversible electroporation ablation procedure using a farawave catheter irrigation of the device stopped.Irrigation during most of the procedure and during preparation had been as expected, but when working on the final vein they lost irrigation while in flower position.Irrigation resumed when the array was changed to basket shape.The procedure was completed with this original catheter with no patient complications.The catheter is expected to be returned for analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FARAWAVE PULSED FIELD ABLATION CATHETER, 35MM
Type of Device
CARDIAC IRREVERSIBLE ELEC TROPORATION SYSTEM CATHETER
Manufacturer (Section D)
FARAPULSE, INC.
3715 haven avenue
suite 110
menlo park CA 94025
Manufacturer (Section G)
AGILE MV INC
3700 st patrick st suite 102
montreal, qc H4E 1 A2
CA   H4E 1A2
Manufacturer Contact
timothy degroot
4100 hamline avenue north
dc a330
saint paul, MN 55112
6515826168
MDR Report Key18932862
MDR Text Key338084508
Report Number2124215-2024-16342
Device Sequence Number1
Product Code QZI
Combination Product (y/n)N
Reporter Country CodeAS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 03/19/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Lot Number0008020937
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/29/2024
Initial Date FDA Received03/19/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/06/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-