• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION ACCOLADE MRI DR; PACEMAKERS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION ACCOLADE MRI DR; PACEMAKERS Back to Search Results
Model Number L311
Device Problems Incorrect, Inadequate or Imprecise Result or Readings (1535); Battery Problem (2885); High Capture Threshold (3266)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/07/2024
Event Type  Injury  
Event Description
It was reported that that patient with this pacemaker device was found in safety mode when the device was programmed for magnetic resonance imaging (mri) mode.Technical services (ts) reviewed and stated that device replacement is needed.There was an increase in power consumption due to increase in pacing output voltage.Subsequently this device was explanted and successfully replaced.No additional patient adverse effects were reported.
 
Manufacturer Narrative
Upon receipt at our post market quality assurance laboratory, this device was thoroughly inspected and analyzed.Device telemetry data confirmed it was operating in safety mode and that bradycardia therapy remained available.Engineers determined that this device was demonstrating behavior consistent with a high internal cell impedance within the battery.The high internal impedance put this device at risk of experiencing transient voltage decreases and associated resets during periods of high-power consumption.When battery voltage drops below a minimum threshold, a system reset is performed.If three system resets occur within a 48-hour period, the device is designed to immediately enter safety mode operation to maintain back-up pacing with pre-defined, non-programmable settings.
 
Event Description
It was reported that that patient with this pacemaker device was found in safety mode when the device was programmed for magnetic resonance imaging (mri) mode.Technical services (ts) reviewed and stated that device replacement is needed.There was an increase in power consumption due to increase in pacing output voltage.Subsequently this device was explanted and successfully replaced.No additional patient adverse effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACCOLADE MRI DR
Type of Device
PACEMAKERS
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key18933140
MDR Text Key338025644
Report Number2124215-2024-16525
Device Sequence Number1
Product Code LWP
UDI-Device Identifier00802526559228
UDI-Public00802526559228
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P150012/S000
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Physician
Remedial Action Notification
Type of Report Initial,Followup
Report Date 05/15/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date04/19/2019
Device Model NumberL311
Device Catalogue NumberL311
Device Lot Number290662
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/07/2024
Initial Date FDA Received03/19/2024
Supplement Dates Manufacturer Received05/14/2024
Supplement Dates FDA Received05/15/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/10/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age75 YR
Patient SexMale
-
-