H10: the device was not received for evaluation; however, the machine was evaluated on-site by a baxter qualified technician.Functional testing included inspection of the condition of the riva, reva, zeva, tava, diva, and byva valve membranes, as well as the degassing chamber and power supply.No defects were noted and the condition was reproduced.A service history review revealed no indication that the parts replaced during servicing caused or contributed to the reported event.The reported condition was verified.The cause of the condition was determined to be the defective ultrafiltration unit which was replaced to address this issue.Should additional relevant information become available, a supplemental report will be submitted.
|