MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION AK 96 MACHINES; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION AK 96 MACHINES; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM Back to Search Results
Catalog Number 114362
Device Problem Inadequate Ultra Filtration (1656)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/01/2024
Event Type  malfunction  
Event Description
It was reported that "too much equipment is dehydrated" during hemodialysis therapy with an ak 96 machine.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Manufacturer Narrative
H10: the device was not received for evaluation; however, the machine was evaluated on-site by a baxter qualified technician.Functional testing included inspection of the condition of the riva, reva, zeva, tava, diva, and byva valve membranes, as well as the degassing chamber and power supply.No defects were noted and the condition was reproduced.A service history review revealed no indication that the parts replaced during servicing caused or contributed to the reported event.The reported condition was verified.The cause of the condition was determined to be the defective ultrafiltration unit which was replaced to address this issue.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AK 96 MACHINES
Type of Device
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - LUND MONITORS
magistratsvagen 16
lund skane lan SE-22 643
SW   SE-22643
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key18933383
MDR Text Key338027420
Report Number9616026-2024-00015
Device Sequence Number1
Product Code KDI
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 05/07/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number114362
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 03/01/2024
Initial Date FDA Received03/19/2024
Supplement Dates Manufacturer Received05/02/2024
Supplement Dates FDA Received05/07/2024
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-