• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON DICKINSON YRINGE 5ML SALINE FILL CHINA SP; SALINE, VASCULAR ACCESS FLUSH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON YRINGE 5ML SALINE FILL CHINA SP; SALINE, VASCULAR ACCESS FLUSH Back to Search Results
Catalog Number 306594
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/04/2024
Event Type  malfunction  
Manufacturer Narrative
Pr (b)(4): initial mdr submission.A follow up mdr will be submitted if additional information, a device evaluation, or a device history review is completed.
 
Event Description
On (b)(6) 2024, a nurse used a prefilled catheter irrigator to flush an indwelling needle catheter for a patient, and found that its outer packaging was damaged and the push rod handle was broken.Then replace the washer with a new one.
 
Manufacturer Narrative
(b)(4) follow up it was reported the outer packaging was damaged and the push rod handle was broken.As a sample was not returned, a thorough sample investigation could not be performed.To aid in the investigation, two photos were provided for evaluation by our quality team.The photos show a syringe in its packaging flow wrap.The syringe plunger rod thumb press and the packaging flow wrap are damaged.No other defects or imperfections were observed.This defect could occur if the syringe was not placed in the correct location on the conveyor inducing the damage.A device history record review was completed for provided material number 306594, lot 3297902.The review did not reveal any detected quality issues during the production of this lot that could have contributed to the reported defect.There were no related quality notifications.All processes and final inspections complied with specification requirements.Verification of the flow wrap process was performed.The settings were correct, and the flow of product was good.To date, there have been no other similar events reported for this lot.Based on the investigation and with the photo sample analysis the symptom reported by the customer is confirmed.
 
Event Description
No additional information received.On (b)(6) 2024, a nurse used a prefilled catheter irrigator to flush an indwelling needle catheter for a patient, and found that its outer packaging was damaged and the push rod handle was broken.Then replace the washer with a new one.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
YRINGE 5ML SALINE FILL CHINA SP
Type of Device
SALINE, VASCULAR ACCESS FLUSH
Manufacturer (Section D)
BECTON DICKINSON
1 becton drive
franklin lakes NJ 07417
Manufacturer (Section G)
BD MEDICAL (BD WEST) MEDICAL SURGICAL
1852 10th avenue
columbus NE 68601
Manufacturer Contact
helen cox (mdr)
75 north fairway drive
vernon hills, IL 60061
8473935694
MDR Report Key18933469
MDR Text Key338660708
Report Number1911916-2024-00196
Device Sequence Number1
Product Code NGT
UDI-Device Identifier30382903065944
UDI-Public(01)30382903065944
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K161552
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 04/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number306594
Device Lot Number3297902
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/13/2024
Initial Date FDA Received03/19/2024
Supplement Dates Manufacturer Received04/26/2024
Supplement Dates FDA Received04/26/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/24/2023
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-