• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ACON LABORATORIES, INC. FLOWFLEX COVID-19 ANTIGEN HOME TEST; ANTIGEN DIAGNOSTIC FOR SARS-COV-2, PRODUCT CODE: QKP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ACON LABORATORIES, INC. FLOWFLEX COVID-19 ANTIGEN HOME TEST; ANTIGEN DIAGNOSTIC FOR SARS-COV-2, PRODUCT CODE: QKP Back to Search Results
Lot Number COV3090006
Device Problem Unable to Obtain Readings (1516)
Patient Problem Insufficient Information (4580)
Event Date 03/14/2024
Event Type  malfunction  
Manufacturer Narrative
The current complaint is pending investigation from our contract manufacturer, and we will provide follow up mdr upon investigation completion such as review of batch records or testing of retention samples.Any additional information received by acon may be provided to fda in a follow up mdr.
 
Event Description
Invalid result (s).Invalid result.The customer stated that their cassette did not produce a result.The customer confirmed that at least 4 drops were added into the sample well.
 
Event Description
Invalid result (s).Invalid result.The customer stated that their cassette did not produce a result.The customer confirmed that at least 4 drops were added into the sample well.
 
Manufacturer Narrative
Batch records for final product manufacture and qc record for cov3090006 were reviewed.No abnormal issue was found in manufacturing process, technical testing and quality control inspection, and the manufacturing process is complied with dmr.Retention samples from cov3090006 were tested and met the qc criteria.The issue was not found from the retained cassettes.The complaint could not be verified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLOWFLEX COVID-19 ANTIGEN HOME TEST
Type of Device
ANTIGEN DIAGNOSTIC FOR SARS-COV-2, PRODUCT CODE: QKP
Manufacturer (Section D)
ACON LABORATORIES, INC.
5850 oberlin drive #340
san diego CA 92121
Manufacturer (Section G)
ACON LABORATORIES, INC.
5850 oberlin drive #340
san diego CA 92121
Manufacturer Contact
qiyi xie
MDR Report Key18933835
MDR Text Key338441123
Report Number2531491-2024-00039
Device Sequence Number1
Product Code QKP
Combination Product (y/n)N
PMA/PMN Number
EUA210494
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Other
Type of Report Followup
Report Date 03/19/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Lot NumberCOV3090006
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received Not provided
Initial Date FDA Received03/19/2024
Supplement Dates Manufacturer Received03/14/2024
Supplement Dates FDA Received04/26/2024
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-