• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: THORATEC SWITZERLAND GMBH CENTRIMAG 2ND GENERATION PRIMARY CONSOLE; Control, pump speed, cardiopulmonary bypass

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

THORATEC SWITZERLAND GMBH CENTRIMAG 2ND GENERATION PRIMARY CONSOLE; Control, pump speed, cardiopulmonary bypass Back to Search Results
Model Number 201-90411
Device Problem Infusion or Flow Problem (2964)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/27/2024
Event Type  malfunction  
Manufacturer Narrative
No further information was provided.A supplemental report will be submitted once the manufacturer¿s investigation is complete.
 
Event Description
It was reported that the patient, who was on a pedimag, had suction events on the evening on (b)(6) 2023.No alarms were triggered.Log files were submitted for review and captured a low flow event showing a drop to 0 liters per minute (lpm), associated with a loss of flow signal.The event was muted or cleared within 5 seconds.The flow dropped to approximately 1.0 lpm (lowest value recorded was 0.9 lpm.) this did not result in an alarm because the alarm threshold was set to 0.5 lpm at the time.There was another low flow showing 0 lpm, again associated with a loss of flow signal.This alarm was also muted/cleared.Again, there was another low flow showing 0 lpm, again associated with a loss of flow signal, which was also muted/cleared.Related manufacturer reference number: 3003306248-2024-00442, 3003306248-2024-00441, 3003306248-2024-00440.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTRIMAG 2ND GENERATION PRIMARY CONSOLE
Type of Device
Control, pump speed, cardiopulmonary bypass
Manufacturer (Section D)
THORATEC SWITZERLAND GMBH
technoparkstrasse 1
zurich CH-80 05
SZ  CH-8005
Manufacturer (Section G)
THORATEC SWITZERLAND GMBH
technoparkstrasse 1
zurich CH-80 05
SZ   CH-8005
Manufacturer Contact
lindsey sallese
6035 stoneridge drive
pleasanton, CA 94588
7818528207
MDR Report Key18935481
MDR Text Key338981456
Report Number3003306248-2024-00439
Device Sequence Number1
Product Code DWA
UDI-Device Identifier07640135140702
UDI-Public07640135140702
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K131179
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Nurse
Type of Report Initial
Report Date 03/19/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number201-90411
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 02/28/2024
Initial Date FDA Received03/19/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-