• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: GYRUS ACMI, INC. SHOCKPULSE LITHOTRIPSY TRANSDUCER.; LITHOTRIPTOR, ULTRASONIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GYRUS ACMI, INC. SHOCKPULSE LITHOTRIPSY TRANSDUCER.; LITHOTRIPTOR, ULTRASONIC Back to Search Results
Model Number SPL-T
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/20/2024
Event Type  malfunction  
Manufacturer Narrative
To date, device has not yet been returned.A supplemental report will be submitted when the investigation is completed or if additional information becomes available.
 
Event Description
It was reported, the shockpulse lithotripsy transducer ¿locking clip is broken¿.The issue was found during preparation and prior to an unspecified diagnostic procedure.There was no report of patient harm associated with the event.
 
Event Description
It was reported, the shockpulse lithotripsy transducer "locking clip is broken".The issue was found during preparation and prior to an unspecified diagnostic procedure.There was no report of patient harm associated with the event.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation and device evaluation.The device was returned to olympus for evaluation and the customer¿s allegation was not confirmed.It was determined that the device met performance specifications.A review of the device history record found no deviations that could have caused or contributed to the issue.Based on the results of the investigation, the definitive root cause of the issue could not be determined.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SHOCKPULSE LITHOTRIPSY TRANSDUCER.
Type of Device
LITHOTRIPTOR, ULTRASONIC
Manufacturer (Section D)
GYRUS ACMI, INC.
9600 louisiana avenue north
brooklyn park MN 55445
Manufacturer (Section G)
GYRUS ACMI, INC.
9600 louisiana avenue north
brooklyn park MN 55445
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18937863
MDR Text Key338085881
Report Number3011050570-2024-10040
Device Sequence Number1
Product Code FEO
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K142428
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 05/10/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberSPL-T
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received03/19/2024
Supplement Dates Manufacturer Received04/11/2024
Supplement Dates FDA Received05/10/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/25/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-