It was reported that this right ventricular (rv) lead exhibited variable r wave amplitude and a few high threshold measurements.Technical services (ts) reviewed noting this patient appeared to be in bigeminy at the beginning of the electrogram (egm) strip however the last four beats were not bigeminy and were declaring loss of capture (loc).Ts could not decipher if this was true loc or if the threshold test was ending early.Ts recommended to the health care professional, in clinic rv threshold testing.The device was currently programmed to left ventricular (lv) only and the presenting rhythm showed lv pacing and capture.This rv lead remains in service.No adverse patient effects were reported.
|