Model Number BF-P60
Device Problem Peeled/Delaminated (1454)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The device evaluation is ongoing.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was observed that during the device evaluation, the bronchofiberscope exhibited that the insertion tube coating had peeling.There were no reports of patient involvement.
 
Manufacturer Narrative
Correction to e1, e2, e3 and g2 with information that was inadvertently missed on the initial medwatch.This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.The device was returned, and an evaluation was completed.A review of device history record and historical complaints analysis was conducted and no deviations that could have caused or contributed to the reported issue were identified.Based on the results of the investigation, it is likely physical stress was applied to insertion section, stress from inferior storage environment (in direct sunlight, at high temperatures, in high humidity or exposed to x-rays and/or ultraviolet-rays, etc.), and chemical stress from reprocessing not recommended by instructions for use (ifu) caused the coating to peel.However, a definitive root cause could not be determined.The following information is stated in the instructions for use (ifu): 3.2 ¿inspection of the endoscope.¿ 5.¿inspect the external surface of the entire insertion tube including the bending section and the distal end for dents, bulges, swelling, peeling, scratching, holes, sagging, transformation, bends, adhesion of foreign body, dropout of parts, any protruding objects or other irregularities.¿ olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OES BRONCHOFIBERSCOPE
Type of Device
BRONCHOFIBERSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18940148
MDR Text Key338568259
Report Number9610595-2024-05901
Device Sequence Number1
Product Code EOQ
UDI-Device Identifier04953170308123
UDI-Public04953170308123
Combination Product (y/n)N
Reporter Country CodeKS
PMA/PMN Number
K023984
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 04/12/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberBF-P60
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/04/2024
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 03/18/2024
Initial Date FDA Received03/20/2024
Supplement Dates Manufacturer Received04/04/2024
Supplement Dates FDA Received04/12/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/16/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-