ST. JUDE MEDICAL BRK-1¿ XS TRANSSEPTAL NEEDLE, 71 CM LENGTH; Trocar Back to Search Results
Model Number G407209
Device Problem Gas/Air Leak (2946)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/19/2024
Event Type  malfunction  
Event Description
During an atrial fibrillation ablation procedure, an air leak occurred.Once the needle was inserted into the patient, when attempting to pull blood back, air entered the system.It seemed the threading between the flush and needle was not tight.The needle was replaced and the procedure continued with no patient consequences.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BRK-1¿ XS TRANSSEPTAL NEEDLE, 71 CM LENGTH
Type of Device
Trocar
Manufacturer (Section D)
ST. JUDE MEDICAL
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS  1897-4050
Manufacturer (Section G)
ST. JUDE MEDICAL
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS   1897-4050
Manufacturer Contact
janna parks
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key18940233
MDR Text Key338126157
Report Number3008452825-2024-00153
Device Sequence Number1
Product Code DRC
UDI-Device Identifier05414734205184
UDI-Public05414734205184
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 03/20/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberG407209
Device Lot Number10084016
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 03/19/2024
Initial Date FDA Received03/20/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/15/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-