• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AIZU OLYMPUS CO., LTD. ENDOEYE FLEX DEFLECTABLE VIDEOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AIZU OLYMPUS CO., LTD. ENDOEYE FLEX DEFLECTABLE VIDEOSCOPE Back to Search Results
Model Number LTF-S190-10
Device Problem Poor Quality Image (1408)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
It was reported the deflectable videoscope had image clouding.The issue occurred during preparation for use.There were no reports of patient harm.
 
Manufacturer Narrative
E1: (b)(6) hospital.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the image clouding was presumed to be from moisture/foreign material invasion in the subject device; however, a definitive root cause could not be identified.The customer may be able to reduce and prevent occurrence of the event by handling device in accordance with the following ifu: operation manual: 3.8 inspection of the endoscopic system: inspection of the endoscopic image.If additional information becomes available at a later date, this report will be supplemented.Olympus will continue to monitor the field performance of this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOEYE FLEX DEFLECTABLE VIDEOSCOPE
Type of Device
FLEX DEFLECTABLE VIDEOSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA   965-8520
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18940358
MDR Text Key338448465
Report Number9610595-2024-05907
Device Sequence Number1
Product Code GCJ
UDI-Device Identifier04953170434662
UDI-Public04953170434662
Combination Product (y/n)N
Reporter Country CodeKS
PMA/PMN Number
K111425
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility
Reporter Occupation Other
Type of Report Initial
Report Date 03/20/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberLTF-S190-10
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/21/2021
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/21/2021
Initial Date FDA Received03/20/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/24/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-