RESPIRONICS, INC. RESPIRONICS; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE Back to Search Results
Model Number V60 V60PLUS VENTILATOR
Device Problem Failure to Sense (1559)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/29/2024
Event Type  malfunction  
Event Description
Philips received a complaint by the field service engineer (fse) on the v60 indicating that while servicing the device, it was observed that the drpt event log displayed error code 110b proximal pressure sensor auto zero failed message during pre-op check.It was reported there was no patient involvement at the time the issue was discovered.Investigation ongoing.
 
Manufacturer Narrative
The manufacturer's field service engineer (fse) was dispatched to the site and confirmed the reported problem.The fse found error code 110b in the logs while working on fco 66.The fse proceeded to run device for 20 minutes and found proximal pressure sensor auto zero failed error populating screen.The fse reviewed the event logs and service manual and found that the device is in need of two solenoid valve replacements.The fse replaced the two solenoid valves to resolve the reported issue.The device passed required performance verification tests per philips standards and was returned to service.The investigation concludes that no further action is required at this time.If additional information is received the complaint file will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESPIRONICS
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
kimberly shelly
1001 murry ridge lane
murrysville, PA 15668
7247330200
MDR Report Key18940914
MDR Text Key338133111
Report Number2518422-2024-14490
Device Sequence Number1
Product Code MNT
UDI-Device Identifier00884838020054
UDI-Public00884838020054
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102985
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberV60 V60PLUS VENTILATOR
Device Catalogue Number1053617
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 02/29/2024
Initial Date FDA Received03/20/2024
Supplement Dates Manufacturer Received04/26/2024
Supplement Dates FDA Received05/06/2024
Date Device Manufactured05/18/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-