Ref.Id: (b)(4).The digitaldiagnost c90 is a stationary x-ray system for general radiographic purposes.As an option, a portable digital flat panel detector (model "skyplate") can be used for image capture.The local field service engineer (fse) went on site and inspected the skyplate detector.He successfully completed gain and pixel calibrations.He tested the image and returned the system back to the customer with full functionality.Finally, the system meets meets the specification for the performed service and is returned to use.The detector dropped out of the users hand and registered as a medium shock (use error).Risk estimation revealed acceptable risk per risk benefit analysis, because actually, there is no feasible technical solution for this kind of ¿use error¿.This issue is further monitored and trended.
|