Model Number M4735A
Device Problems Failure to Discharge (1169); Failure of Device to Self-Test (2937)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/20/2023
Event Type  malfunction  
Event Description
It was reported to philips that the device failed the therapy delivery test.There was no patient involvement.The efficia dfm100 defibrillator, model# 866199, is substantially similar to the heartstart xl+ defibrillator (model # 861290) and will be reported in the united states under device model # 861290.
 
Manufacturer Narrative
The device was evaluated by customer only.Customer confirmed that the power pca was damaged.However, since the device is out of warranty, customer decided to repair it by themselves.Based on the information available, the cause of the reported problem was defective power pca.The reported problem was confirmed.The device was operational after repairs were completed by customer.Device passed all required tests, and it remains at customer site.The investigation concludes that no further action is required at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEART START XL
Type of Device
DEFIBRILLATOR
Manufacturer (Section D)
PHILIPS GOLDWAY (SHENZHEN) INDUSTRIAL INC.
no.2 keji north 3rd road
nanshan district
shenzhen
CH 
Manufacturer (Section G)
PHILIPS GOLDWAY (SHENZHEN) INDUSTRIAL INC.
no.2 keji north 3rd road
nanshan district
shenzhen
CH  
Manufacturer Contact
feng she
no.2 keji north 3rd road
nanshan district
shenzhen 
CH  
7552698099
MDR Report Key18942224
MDR Text Key338144326
Report Number3030677-2024-01049
Device Sequence Number1
Product Code MKJ
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K021453
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberM4735A
Device Catalogue NumberM4735A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 02/23/2024
Initial Date FDA Received03/20/2024
Supplement Dates Manufacturer Received03/22/2024
Supplement Dates FDA Received03/22/2024
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-