It was reported to philips that the device failed the therapy delivery test.There was no patient involvement.The efficia dfm100 defibrillator, model# 866199, is substantially similar to the heartstart xl+ defibrillator (model # 861290) and will be reported in the united states under device model # 861290.
|
The device was evaluated by customer only.Customer confirmed that the power pca was damaged.However, since the device is out of warranty, customer decided to repair it by themselves.Based on the information available, the cause of the reported problem was defective power pca.The reported problem was confirmed.The device was operational after repairs were completed by customer.Device passed all required tests, and it remains at customer site.The investigation concludes that no further action is required at this time.
|