Model Number V60 V60PLUS VENTILATOR
Device Problem Display Difficult to Read (1181)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/18/2024
Event Type  malfunction  
Event Description
Philips received a complaint by the customer on the v60, indicating that the customer was unable to read the display due to the backlight being out.It was reported there was no patient involvement at the time the issue was discovered.It was reported that the customer was unable to read the display due to the backlight being out.The remote service engineer (rse) informed the customer of the contract status and the customer requested onsite service.Device evaluation and repair are pending.
 
Manufacturer Narrative
H10: a field service engineer (fse) went onsite and confirmed the reported issue.The fse then replaced the light crystal display (lcd) and user interface (ui) printed circuit board assembly (pcba) to resolve the reported issue.The fse replaced the lcd and ui pcba to resolve the reported issue.The device passed required performance verification tests per philips standards and was returned to service.The investigation concludes that no further action is required at this time.If additional information is received the complaint file will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESPIRONICS
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
kimberly shelly
1001 murry ridge lane
murrysville, PA 15668
7247330200
MDR Report Key18943685
MDR Text Key338209418
Report Number2518422-2024-14542
Device Sequence Number1
Product Code MNT
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102985
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberV60 V60PLUS VENTILATOR
Device Catalogue Number1053617
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/18/2024
Initial Date FDA Received03/20/2024
Supplement Dates Manufacturer Received03/28/2024
Supplement Dates FDA Received04/02/2024
Date Device Manufactured09/17/2009
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-