Model Number V60
Device Problem Failure to Power Up (1476)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/04/2024
Event Type  malfunction  
Event Description
Philips received a complaint from the customer on the v60 indicating that the device will not operate correctly when it is powered on; the display screen is blank intermittently, and it alarms.There was no patient involvement at the time the issue was discovered.There was no report of patient or user harm.The customer evaluated the device with the assistance of the remote service engineer (rse) and confirmed the reported problem.The biomedical engineer customer indicated that the v60 ventilator has the new 4th generation power management board installed.The rse advised the customer that the display appears to be functioning correctly and the problem could be the power management board.The rse provided the customer with the part number for the power management board.
 
Manufacturer Narrative
Multiple attempts have been made to try to obtain further information about this case, but no response received from the customer.The complaint will be processed for closure.If additional information becomes available at a later date, the complaint will be reopened, and a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESPIRONICS
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
kimberly shelly
1001 murry ridge lane
murrysville, PA 15668
7247330200
MDR Report Key18946371
MDR Text Key338239404
Report Number2518422-2024-14584
Device Sequence Number1
Product Code MNT
UDI-Device Identifier00884838020054
UDI-Public00884838020054
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102985
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberV60
Device Catalogue Number1053617
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/04/2024
Initial Date FDA Received03/20/2024
Supplement Dates Manufacturer Received05/03/2024
Supplement Dates FDA Received05/08/2024
Date Device Manufactured02/20/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-