Model Number HAMILTON-C1 |
Device Problem
Defective Alarm (1014)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/28/2022 |
Event Type
malfunction
|
Event Description
|
The report from hospital say: on (b)(6) 2022, the ventilator gave a technical defect alarm during pre-use inspection, showing alarm code "technical event: 285002".Hamilton-c1 made in jun.18, 2016.
|
|
Manufacturer Narrative
|
The complaint has been reopened and reviewed according to fda form 483 inspectional observation ems #2, eobs2 from the fda inspection conducted between july 17 to july 21, 2022 at the ems and bonaduz sites.A detailed investigation was performed by an expert from the technical service: no further investigation or correction will be performed except those mentioned above.In future hamilton medical ag will report an event similar to this issue as it will be deemed a reportable event.The allegation in this complaint was confirmed to be a complaint.With this investigation it has been confirmed that the device failed to meet its specifications at the time of the event while the ventilator was prepared for treatment.The root cause could not be determined as the customer hired a third-party maintenance company for repair but was not informed about the specific defect and the repair details.After repair the ventilator went back into service.The hospital was recommended to find a maintenance company authorized by hamilton for repair.There was no patient or user harm reported.
|
|