Model Number CR2 |
Device Problem
Audible Prompt/Feedback Problem (4020)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/27/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
The customer received a replacement device.Stryker continues to investigate the reported failure and will submit a supplemental report on this event to the fda as provided by 21 cfr 803.56.
|
|
Event Description
|
The customer contacted stryker to report that their device did not provide any voice prompts when the device was powered on.Without voice prompts, a lay user would not receive the audible instructions on how to properly use the device on a patient.There was no patient use associated with the reported event.
|
|
Manufacturer Narrative
|
Stryker evaluated the customer's device verified the reported issue and determined it was caused by a depleted battery.A cause for the depleted battery could not be determined.
|
|
Event Description
|
The customer contacted stryker to report that their device did not provide any voice prompts when the device was powered on.Without voice prompts, a lay user would not receive the audible instructions on how to properly use the device on a patient.There was no patient use associated with the reported event.
|
|