Catalog Number SBI05015013P
Device Problem Packaging Problem (3007)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/19/2024
Event Type  malfunction  
Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event Description
It was reported that an attempt was made to use a impact admiral to treat a calcified plaque lesion in the mid superficial femoral artery. degree of tortuosity was little. degree of calcification was moderate.Lesion stenosis was 80%.Device was prepped per ifu with issues noted.It is reported there was a packaging issue.The sterile packaging bag was damaged and device could not be used, there was a gap at the opening of the sterile bag of the product.No obvious damage was observed in the outer box, but there was slight squeeze deformation.The device was not used in the patient.The drug-coated balloon use was cancelled and the procedure was completed.No patient injury reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IN.PACT ADMIRAL PACLITAXEL-ELUTING PTA BALLOON CATHETER
Type of Device
DRUG-ELUTING PERIPHERAL TRANSLUMINAL ANGIOPLAS
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
EI 
Manufacturer (Section G)
MEDTRONIC IRELAND
parkmore business park west
galway
EI  
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key18948759
MDR Text Key338560053
Report Number9612164-2024-01374
Device Sequence Number1
Product Code ONU
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
P140010
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 03/21/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberSBI05015013P
Device Lot Number0011476573
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/15/2024
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/26/2024
Initial Date FDA Received03/21/2024
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age72 YR
Patient SexFemale
-
-