Model Number D-AVHD-DF16
Device Problems Device Contamination with Body Fluid (2317); Obstruction of Flow (2423); Patient Device Interaction Problem (4001)
Patient Problem Thrombosis/Thrombus (4440)
Event Date 03/13/2024
Event Type  Injury  
Event Description
During the procedure, the internal lumen of the catheter was obstructed with a blood clot and could not be flushed.The catheter was used initially to collect geometry and map (approximately 10/15 minutes).The mapping catheter was then replaced with an ablation catheter, leaving the hd grid on the table for ablation (about 30 minutes).The catheter was then tested to reuse for the final remap, and the clot was noted.Another hd grid was used to complete the procedure with no consequences to the patient.
 
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.Based on the information received, the cause of the reported obstruction and blood clot remains unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVISOR¿ HD GRID MAPPING CATHETER, SENSOR ENABLED¿
Type of Device
CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING
Manufacturer (Section D)
ST. JUDE MEDICAL
5050 nathan lane north
plymouth MN 55442
Manufacturer (Section G)
ST. JUDE MEDICAL
5050 nathan lane north
plymouth MN 55442
Manufacturer Contact
janna parks
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key18948849
MDR Text Key338230257
Report Number3005334138-2024-00098
Device Sequence Number1
Product Code DRF
UDI-Device Identifier05415067028198
UDI-Public05415067028198
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
K172393, K20
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/12/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberD-AVHD-DF16
Device Lot Number10047890
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 03/13/2024
Initial Date FDA Received03/21/2024
Supplement Dates Manufacturer Received04/09/2024
Supplement Dates FDA Received04/12/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured10/23/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age74 YR
Patient SexMale
-
-