RESPIRONICS, INC. RESPIRONICS; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
|
Back to Search Results |
|
Model Number V60 V60PLUS VENTILATOR |
Device Problem
Communication or Transmission Problem (2896)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/04/2024 |
Event Type
malfunction
|
Event Description
|
Philips received a complaint by the customer on the v60 indicating that the device is getting serial peripheral interface bus (spi bus) failure (1113-barometer calibration data error, 110f-o2 flow sensor calibration data error, 110e-air flow sensor calibration data error, 1110-o2 pressure sensor calibration data error, 1107-proximal pressure sensor calibration data error, 1106-machine pressure sensor calibration data error) failure message.It was reported there was no patient involvement at the time the issue was discovered.The investigation is ongoing.
|
|
Manufacturer Narrative
|
Multiple attempts have been made to try to obtain further information about this case, but no response received from the customer.The investigation concludes that no further action is required at this time.
|
|