MAUDE Adverse Event Report: MEDTRONIC EXTENDED; UNO EWIS BLUE 60/9 HCAP 3-PK INT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC EXTENDED; UNO EWIS BLUE 60/9 HCAP 3-PK INT Back to Search Results
Lot Number UNKNOWN
Device Problem Material Twisted/Bent (2981)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
Unomedical reference number (b)(4).Event occurred in the united states.On (b)(6) 2024, the patient's father reported that his child faced a kinked cannula leading to high blood glucose level.The site location was patient's leg, and the infusion set was used for one day.Reportedly, the patient lost four infusion set as they were repeatedly getting insulin flow block alarm which resolved by changing the infusion set.Unomedical do not see bent/kinking as being related to human factors, but rather as a training issue including correct choices of insertion sites and infusion sets and cannula length.Furthermore, the soft cannula is a flexible material that during use and upon removal can bend slightly.No further information was available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDTRONIC EXTENDED
Type of Device
UNO EWIS BLUE 60/9 HCAP 3-PK INT
Manufacturer Contact
aaholmvej 1-3
osted
lejre, 4320
MDR Report Key18949968
MDR Text Key339323613
Report Number8021545-2024-00132
Device Sequence Number1
Product Code FPA
UDI-Device Identifier05705244023284
UDI-Public05705244023284
Combination Product (y/n)N
PMA/PMN Number
K210544
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Distributor
Reporter Occupation Other
Type of Report Initial
Report Date 03/21/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Lot NumberUNKNOWN
Initial Date Manufacturer Received 03/14/2024
Initial Date FDA Received03/21/2024
Patient Sequence Number1
-
-