Catalog Number 381812 |
Device Problem
Retraction Problem (1536)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/27/2024 |
Event Type
malfunction
|
Event Description
|
It was reported that bd insyte autoguard needle did not retract.The following information was provided by the initial reporter, translated from chinese to english: safety mechanism is not work, the steel needle cannot be retracted.
|
|
Manufacturer Narrative
|
H.3.A follow up mdr will be submitted if additional information, a device evaluation, or a device history review is completed.
|
|
Manufacturer Narrative
|
Investigation results: the complaint that the needle could not be retracted was confirmed from the photograph that was provided for investigation.The photo showed an insyte autoguard needle and shield without the iv catheter.The needle remained fully extended from the grip.Without the physical sample, the root cause of the damage could not be determined.A review of the inspection records and quality/manufacturing controls for the implicated lot indicated no issues with the manufacturing process.The complaint has been documented and will continue to be monitored as part of ongoing efforts to identify potential manufacturing related issues.
|
|
Event Description
|
No additional information.
|
|