MAUDE Adverse Event Report: BECTON DICKINSON INFUSION THERAPY SYSTEMS INC. BD INSYTE AUTOGUARD; PERIPHERAL IV CATHETERS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON INFUSION THERAPY SYSTEMS INC. BD INSYTE AUTOGUARD; PERIPHERAL IV CATHETERS Back to Search Results
Catalog Number 381812
Device Problem Retraction Problem (1536)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/27/2024
Event Type  malfunction  
Event Description
It was reported that bd insyte autoguard needle did not retract.The following information was provided by the initial reporter, translated from chinese to english: safety mechanism is not work, the steel needle cannot be retracted.
 
Manufacturer Narrative
H.3.A follow up mdr will be submitted if additional information, a device evaluation, or a device history review is completed.
 
Manufacturer Narrative
Investigation results: the complaint that the needle could not be retracted was confirmed from the photograph that was provided for investigation.The photo showed an insyte autoguard needle and shield without the iv catheter.The needle remained fully extended from the grip.Without the physical sample, the root cause of the damage could not be determined.A review of the inspection records and quality/manufacturing controls for the implicated lot indicated no issues with the manufacturing process.The complaint has been documented and will continue to be monitored as part of ongoing efforts to identify potential manufacturing related issues.
 
Event Description
No additional information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD INSYTE AUTOGUARD
Type of Device
PERIPHERAL IV CATHETERS
Manufacturer (Section D)
BECTON DICKINSON INFUSION THERAPY SYSTEMS INC.
9450 south state street
sandy UT 84070
Manufacturer (Section G)
BECTON DICKINSON INFUSION THERAPY SYSTEMS INC.
9450 south state street
sandy UT 84070
Manufacturer Contact
helen cox (mdr)
75 north fairway drive
vernon hills, IL 60061
8473935694
MDR Report Key18950559
MDR Text Key339247645
Report Number1710034-2024-00210
Device Sequence Number1
Product Code FOZ
UDI-Device Identifier00382903818129
UDI-Public(01)00382903818129
Combination Product (y/n)N
Reporter Country CodeTW
PMA/PMN Number
K201075
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 04/18/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number381812
Device Lot Number3198984
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/28/2024
Initial Date FDA Received03/21/2024
Supplement Dates Manufacturer Received04/18/2024
Supplement Dates FDA Received04/29/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/27/2023
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-