• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC, INC. OPEN PIVOT AP360 HEART VALVE; HEART-VALVE, MECHANICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC, INC. OPEN PIVOT AP360 HEART VALVE; HEART-VALVE, MECHANICAL Back to Search Results
Model Number 505DM26
Device Problems Inadequacy of Device Shape and/or Size (1583); Incomplete Coaptation (2507)
Patient Problem Insufficient Information (4580)
Event Date 03/09/2024
Event Type  Injury  
Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event Description
Medtronic received information that during the implant of this 26mm aortic mechanical valve, it was explanted and replaced with a 24mm mechanical valve of the same model.The reason for the replacement was reported as limited leaflet mobility.It was stated that leaflet motion was not tested with the blue actuator during the implant procedure.It was also stated that the valve was rotated within the annulus in an attempt to obtain appropriate leaflet motion.It was reported that the physician thought that this was a case of patient prosthesis mismatch and that the valve selection was too large and affected leaflet rotation.No additional adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPEN PIVOT AP360 HEART VALVE
Type of Device
HEART-VALVE, MECHANICAL
Manufacturer (Section D)
MEDTRONIC, INC.
3800 annapolis lane
minneapolis MN 55447
Manufacturer (Section G)
MEDTRONIC, INC.
3800 annapolis lane
minneapolis MN 55447
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key18951118
MDR Text Key338258234
Report Number3008592544-2024-00018
Device Sequence Number1
Product Code LWQ
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
P990046
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 03/21/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number505DM26
Device Catalogue Number505DM26
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/11/2024
Initial Date FDA Received03/21/2024
Date Device Manufactured05/19/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-