Model Number GIF-HQ290
Device Problem Device Reprocessing Problem (1091)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/06/2024
Event Type  malfunction  
Manufacturer Narrative
The device evaluation is ongoing.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was observed that during the device evaluation, there was white contamination evident in the air and water channel of the videoscope.There were no reports of patient involvement.
 
Manufacturer Narrative
This supplemental report is being submitted to provide the results of the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 8 years since the subject device was manufactured.Based on the results of the investigation, the foreign material was whitish in color, but its exact composition could not be identified.It is possible that the foreign material was not removed since the reprocessing was not conducted properly due to leakage from the scope connector.The event can be detected and prevented in accordance with the following ifu: ifu states that detection method in gif/cf/pcf-290 series operation manual chapter 3 preparation and inspection.Ifu states that preventive measure in gif/cf/pcf-290 series operation manual 3.8 inspection of the endoscopic system: nothing other than sterile water should be used for air/water feeding.No additives should be put into the sterile water.Non-sterile water may cause patient cross-contamination and/or infection.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS LUCERA ELITE GASTROINTESTINAL VIDEOSCOPE
Type of Device
GASTROINTESTINAL VIDEOSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18951704
MDR Text Key338263409
Report Number9610595-2024-06042
Device Sequence Number1
Product Code FDS
Combination Product (y/n)N
Reporter Country CodeUK
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 04/08/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberGIF-HQ290
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/11/2024
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 03/19/2024
Initial Date FDA Received03/21/2024
Supplement Dates Manufacturer Received04/02/2024
Supplement Dates FDA Received04/08/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/12/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-