• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AIZU OLYMPUS CO., LTD. FORCEPS/IRRIGATION PLUG (ISOLATED TYPE); FORCEPS/ IRRIGATION PLUG

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AIZU OLYMPUS CO., LTD. FORCEPS/IRRIGATION PLUG (ISOLATED TYPE); FORCEPS/ IRRIGATION PLUG Back to Search Results
Model Number MAJ-891
Device Problem Contamination (1120)
Patient Problems Fever (1858); Unspecified Infection (1930)
Event Type  Injury  
Manufacturer Narrative
The device evaluation is ongoing.Should additional relevant information become available, a supplemental report will be submitted.It was reported there were no problems with equipment handling, and there was no request for facility training.The customer confirmed the device was properly reprocessed in accordance with the instruction manual.The main body, cock, forceps plug, and tightening knob (cap) were all disassembled and were cleaned and disinfected in accordance with the instruction manual.
 
Event Description
It was reported that there was an increase in the number of infections about a year ago and the director felt it was 10-20 people.Some patients were admitted to other hospitals with fever during the night and on saturdays and sundays.A culture test on the device was not performed.The event was reported after use of the forceps/irrigation plug in association with the cysto-nephro videoscope.Additional information regarding patient symptoms, diagnosis, and treatment was requested, but the customer declined to provide it.This report is for patient 19 of 20.
 
Manufacturer Narrative
This report is being submitted to correct the legal manufacturer¿s contact information and facility registration number.The facility registration number is 9610595.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FORCEPS/IRRIGATION PLUG (ISOLATED TYPE)
Type of Device
FORCEPS/ IRRIGATION PLUG
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
odakura, nishigo-mura,
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
odakura, nishigo-mura,
aizuwakamatsu-shi, fukushima 965-8 520
JA   965-8520
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18952184
MDR Text Key338302437
Report Number3002808148-2024-02707
Device Sequence Number1
Product Code FGB
UDI-Device Identifier04953170063114
UDI-Public04953170063114
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K912120
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Distributor
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 05/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMAJ-891
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/04/2024
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 02/26/2024
Initial Date FDA Received03/21/2024
Supplement Dates Manufacturer Received05/22/2024
Supplement Dates FDA Received05/22/2024
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
CYF-VHA (SN (B)(6)).
Patient Outcome(s) Hospitalization;
-
-