Catalog Number CDS0702-XTW |
Device Problem
Material Separation (1562)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/29/2024 |
Event Type
malfunction
|
Event Description
|
It was reported that during device preparation, the clip delivery system (cds) opened from its packaging (first clip to be used in the case).No damage was evident on the packaging.Upon opening the packaging, a screw was loose in the plastic packaging.The screw was loose from the dc handle, the second from the left under the gripper lever.No preparation steps took place prior to this.The cds was therefore immediately deemed faulty and was not used.A new cds was immediately opened and prepared.There was no significant time delay and the rest of the case was completed without issue.
|
|
Manufacturer Narrative
|
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
|
|
Manufacturer Narrative
|
All available information was investigated, and the reported loose dc handle screw was confirmed via returned device analysis.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no similar complaints reported from this lot.Based on available information and the returned device analysis, the investigation determined the loose dc handle screw to be related to a potential product quality issue.This complaint is within the scope of an exception escalation as the complaint description and device code match the specific issue described in the exception.Therefore, this incident was added to exception (issue) 130612 for further investigation.The issue is being addressed per internal operating procedures.Abbott will continue to trend the performance of these devices.
|
|
Search Alerts/Recalls
|