It was reported that a pre-filled syringe detached from an iv when used for flushing and drawing back blood.No serious injury or adverse impact to a patient or to a user was originally reported.To date, no information has been received to indicate that a user or a patient experienced a death, serious injury, medical intervention, follow-up care, or other adverse health impact associated with the reported problem/issue.Samples of the syringe and the needleless connectors used at the reporting facility were returned for evaluation.Testing of samples concluded that when the syringe is fully engaged onto the needleless connector, no disconnection occurred.However, it was confirmed that disconnection may occur if the syringe is not fully engaged with the needleless connector.The root cause was determined to be the result of improper connection and use of the syringe as instructions for use clearly state the devices if "for iv flush only." in an abundance of caution, and in response to an fda 483 issued for cfn (b)(4) on 22-jan-2024, this medwatch is being filed.If additional relevant information becomes available a supplemental medwatch will be filed.
|