AIZU OLYMPUS CO., LTD. EVIS EXERA II GASTROINTESTINAL VIDEOSCOPE Back to Search Results
Model Number GIF-H180J
Device Problem Device Reprocessing Problem (1091)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The device evaluation is ongoing.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
During the device evaluation, the gastrointestinal videoscope exhibited a white foreign material in the auxiliary jet channel.There were no reports of patient involvement.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It is unknown if the reprocessing steps at the material residue point were deviated from the instruction manual.Based on the results of the investigation, it is likely that the foreign material remained due to insufficient reprocessing.The foreign material was unable to be identified.The event can be prevented by following the instructions for use (ifu): ifu sates the detection method in evis exera ii gif/cf/pcf type 180 series operation manual chapter 3 preparation and inspection.Ifu states the preventive measures in evis exera ii gif/cf/pcf type 180 series reprocessing manual chapter 5 reprocessing the endoscope (and related reprocessing accessories).Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA II GASTROINTESTINAL VIDEOSCOPE
Type of Device
GASTROINTESTINAL VIDEOSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18955432
MDR Text Key338441101
Report Number9610595-2024-06073
Device Sequence Number1
Product Code FDS
UDI-Device Identifier04953170307539
UDI-Public04953170307539
Combination Product (y/n)N
Reporter Country CodeAU
PMA/PMN Number
K051645
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 04/11/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberGIF-H180J
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/07/2024
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 03/14/2024
Initial Date FDA Received03/21/2024
Supplement Dates Manufacturer Received04/10/2024
Supplement Dates FDA Received04/11/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/10/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-