Model Number 96102
Device Problem Application Program Freezes, Becomes Nonfunctional (4031)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/20/2024
Event Type  malfunction  
Manufacturer Narrative
A spacelabs field service engineer (fse) evaluated the reported xhibit central station where they were able to verify the reported issue.It was found the the xhibit central station had lost its license.The fse continued to reload the license and after observing proper device operation through functional and performance testing, the device was returned to the customer for use.While reloading the software resolved the reported issue, the cause of the reported issue could not be determined.
 
Event Description
The customer reported that following a reboot of the xhibit central station, the license was lost, which prevented the clinical staff from further monitoring their patients.There was no patient or user harm associated with this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XHIBIT CENTRAL STATION
Type of Device
XHIBIT CENTRAL
Manufacturer (Section D)
SPACELABS HEALTHCARE
35301 se center st
snoqualmie WA 98065
Manufacturer (Section G)
SPACELABS HEALTHCARE
35301 se center st
snoqualmie WA 98065
Manufacturer Contact
whitney snyder
35301 se center st
snoqualmie, WA 98065
MDR Report Key18956611
MDR Text Key338752627
Report Number3010157426-2024-00044
Device Sequence Number1
Product Code MHX
UDI-Device Identifier10841522127069
UDI-Public011084152212706911210503
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K122146
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 03/20/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number96102
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/20/2024
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/20/2024
Initial Date FDA Received03/21/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/05/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-